CleanSpace

Disinfection

Choose a disinfection agent that can be used on the sensitive equipment. See information below on recommended agents or the guide to compatible agents. The user must determine, by suitable testing, the correct application of the agent on the CleanSpace product for their particular application.

Before disinfection, remove used filters and insert the CleanSpace Cleaning & Storage Plug (CS3011) and clean the respirator thoroughly to remove foreign material (follow the steps in Cleaning the Respirator).

Work in a clean environment and ensure good hand hygiene. Wear personal protective equipment (PPE) including nitrile or vinyl gloves during the cleaning and sanitising process. Ensure adequate ventilation when using disinfecting agents.

SANITIZING AND DISINFECTION AGENTS

NOTE: Failure to remove foreign material before disinfection, will make the process ineffective.

Use a sanitising or disinfection agent that is appropriate for the application and approved by the facility safety manager. Read the agent’s user instructions: usability, application, dilution ratio and contact time.

  • Facepiece/masks, neck supports and harness (NOT POWER UNIT): Soak and rinse in clean water. Air dry the parts before storage
  • Power Unit: Wipe with a soft clean cloth dampened with the agent and then remove agent with a clean cloth. Air dry the respirator before storage

Store sanitised or disinfected respiratory equipment in a clean, sealable, impermeable bag away from contamination, dust and water.

Table 1. Recommended Sanitizing & Disinfection Agents:

Klercide™ Sporicidal Chlorine and Apesin AP3 5% chlorine solution or solution containing 5000ppm available free chlorine.
Klerwipe™ Sporicidal Low Residue Peroxide
Klercide™ 70 Isopropyl alcohol
Miltons (a diluted bleach based agent) can be used. Bleach based agents are generally not recommended
Incidin Plus (Ecolab) Aldehyde-free and without QAV glucoprotamine Active Agent: glucoprotamine
Dismozon (Hartman) Low residue and broad material compatibility. Broad spectrum activity against bacteria, yeast, fungi, spores and virus. Active Agent: Magnesium Monoperoxyphthalate hexahydrate 958 mg/g
Germex™ B-12 Quaternary ammonium compounds, PH 7.3 at 0.75% solution

IMPORTANT: Dispose of used filters according to your safety policy and local regulations. CleanSpace filters and cartridges cannot be cleaned or disinfected.

IMPORTANT: Ensure equipment is thoroughly rinsed with clean water and free of AGENTS before use. Air-dry the power unit in a clean environment before use or storage. After cleaning the power unit, insert the CleanSpace® Cleaning & Storage Plug Set (CS3011) to prevent contaminates entering the respirator when not in use.

CHARGING AND STORAGE
CARE

This information should be reviewed thoroughly by your HSE and infection control managers prior to selecting the disinfecting product for your equipment and specific application. Please read IMPORTANT NOTES ON CLEANING below. Contact our technical team for assistance.

Get In Touch

SPEAK TO OUR CUSTOMER SERVICE TEAM

Contact Us > >